SPECIFICATIONS OF ALL SPR NAVI™ INSTRUMENTS

Technique Multi-Parametric Surface Plasmon Resonance (MP-SPR)
Measuring principle True goniometric SPR with rotating laser (true angluar resolution 0.001o)
Wavelength of light 670 nm (other wavelengths and set-ups such as several lasers per channel available by request)
SPR angle range 40o - 78o
Prism coating Optical gel index matching (no need for messy oil operation between measurements)
Sensor slides  12 x 20mm, gold coated, superior adhesion (other coatings available as option) 
Flow cell

Standard: Easily exchangeable, 2 channels, PDMS material, 100µ cell height
Other: Electrochemistry, Gas, High-resistance (PEEK)

Flow cell volume 1 µl as standard
Sample volume Min. 20 µl (or less through partial loopfill)
Refractive index (RI) range 1.00 - 1.40 (Bulk)
Dynamic range 38 000 mdeg
Noise level 1 µRIU (1 x 10-6 of Refractive Index Unit, a physical measure) - corresponds to 1 pg/mm² surface coverage of a typical protein
Sensitivity

1μRIU - corresponds to 50 ppm change in reflectance signal in the Fixed Angle mode; and to 0.15 mdeg change in SPR peak angular position in the Angular Scan mode

Fastest sampling rate from 4 ms
Temperature control 15-45oC  (ambient +20/-7oC)
Environment and interface SPR Navi control and data viewer software compatible with MS Windows® XP / VISTA / 7 (however, does not support Starter editions used in netbooks), USB 2.0
Power requirements 100W, 100-240V, 50/60Hz

FURTHER SPECIFICATIONS OF SPR NAVI™ 220A

SPR Navi 220A

Liquid handling

 Integrated Autosampler

  • Fully configurable program
  • Sample capacity: 1 microtiter plate, 48 vials (1.5 mL, optional), or 12 vials (10 mL, optional)
  • Loop volume: 1 - 5000 μL programmable, 10 mL loop optional, exchangeable
  • Dispenser syringe: 500 μL standard (250, 1000 and 2500 μL optional
  • Injection modes: Full loop, partial loopfill and μL pickup mode, PASA™ (pressure-assisted sample aspiration)

 

Integrated Syringe pump

  •  2-channel, flowrates 0.831 nL/min - 150.5mL/min, syringe sizes 10 L - 140 mL

Integrated Degasser 

  • 2-channel degasser
  • Internal volume only 480 µl
  • Eliminates baseline fluctuations, advanced error and leak checking functions
  • Wetted parts: PEEK and Systec AF

Dimensions

 62 x 41 x 47 cm (24.5" x 16" x 18.5")

Weight

 

FURTHER SPECIFICATIONS OF SPR NAVI™ 200

SPR Navi 200  Liquid handling  Integrated 2-channel peristaltic pump, software controlled
Sample loading and injection  12-port injector
Syringe pump
Optional 
 2-channel, flowrates 0.831 nL/min - 150.5mL/min, syringe sizes 10 L - 140 mL
Autosampler
Optional 
  •  Fully configurable program
  • Sample capacity: 2 microtiter plates, 2x 48 vials (1.5 mL), or 2x 12 vials (10 mL, optional)
  • Loop volume: 1 - 5000 μL programmable, 10 mL loop optional, exchangeable
  • Dispenser syringe: 500 μL standard (250, 1000 and 2500 μL optional
  • Injection modes: Full loop, partial loopfill and μL pickup mode, PASA™ (pressure-assisted sample aspiration)
  • More info

Degasser
Optional 

 2-channel degasser, internal volume only 480 µl, eliminates baseline fluctuations, advanced error and leak checking functions, wetted parts: PEEK and Systec AF
Dimensions  38 x 32 x 31 cm (15" x 12.5" x 12.4") 
Weight  10kg (22 lbs)

Note on Conversions

1 µRIU (1 x 10-6 of Refractive Index Unit, a physical measure) corresponds to 1 RU (Resonance Unit, an arbitrary unit used by competitors) based on (1)

References:

(1) R.B.M. Schasfoort and A.J. Tudos, Handbook of Surface Plasmon Resonance, RSC Publishing, ISBN 978-0-85404-267-8, pp. 44